Can he tell when a sensor reading should not be trusted?
I set the criteria a wearable signal must pass before anyone builds advice on it, and I say plainly where a sensor stops being reliable. I have tested simple sensors against reference devices in the lab, and I have worked with the reference measures in large cohorts.
How I work:
Evidence
Written standard
Evaluation
A specific fix
Guidance people keep following
For a sensor, evaluation means validation: does the reading agree with a reference device, and does it follow real change in the same person?
The evidence
Finger and toe sensors
We compared a simple photoplethysmography (PPG) sensor with a reference device in 30 adults, at rest and after a tilt test. Finger readings agreed well overall but did not follow change. Toe readings followed change well [1].
Muscle oxygen sensors
We tested two near-infrared spectroscopy (NIRS) methods in healthy men and men with type 2 diabetes on three days. Both were reproducible, but the cheaper method read perfusion higher in diabetes [2].
Reference measures at scale
In 4,183 older adults (ARIC study), a ratio of two arterial stiffness readings was linked with more heart and vascular conditions than the standard measure, and did not depend on blood pressure [3, 4].
Signals in products
In industry, I set the criteria each watch signal must pass with engineering and data science teams, including signal quality and timing calibration checks. I work across accelerometry, PPG, ECG, and heart rate variability.
What a sensor can and cannot support
A one-page summary from our published validation work. Each row states a use the data support and a use they do not.
#
Measure
Can support
Cannot support yet
Source
1
PPG, heart to finger
A single resting estimate of arterial stiffness (overall agreement ICC 0.77)
Tracking change in one person (ICC 0.10)
[1]
2
PPG, heart to toe
Tracking an acute change in one person (ICC 0.81)
Swapping in for the reference reading (overall ICC 0.50)
[1]
3
Any PPG measure timed from the heartbeat
Relative comparisons when timing is handled the same way
A pure stiffness value: the time includes the heart’s pre-ejection period, which was not controlled
[1]
4
Continuous-wave NIRS
Muscle blood flow and oxygen use, at rest and in exercise
Perfusion in type 2 diabetes, where it read higher than frequency-domain NIRS
[2]
5
Aortic-femoral stiffness gradient
A marker that does not depend on blood pressure, in older adults
Clinical decisions, until further studies confirm its use
[3, 4]
ICC: intra-class correlation, where 1 is perfect agreement. Source: the published papers listed below.
Where this fails: a sensor can agree with a reference at rest and still miss change. Our own finger results showed exactly that, so the paper limited its claim to the lower limb.
Sample: a validation checklist for a wrist sensor
Illustrative sample
Illustrative sample. Synthetic data, not from any employer or client. A generic checklist, built from the published work on this page.
#
Check
Pass rule, set before any data are collected
1
Intended use
The decision the reading will support is written down first
2
Reference and protocol
The reference device, positions, and timing are named in advance
3
Signal quality
Poor segments (motion, loose contact) are removed by a fixed rule, and the share removed is reported
4
Agreement
Bias and limits of agreement fall inside limits chosen for the intended use
5
Change
The sensor follows a known change (posture, exercise), not just resting values
6
Timing
Any measure timed from the heartbeat accounts for the pre-ejection period
7
Who was tested
Age, sex, skin tone, body size, and health conditions are reported; untested groups are named
8
Repeatability
Day-to-day reliability is reported with its typical error
A signal is used only for the purposes its checks support.
Sample: an agreement plot, with the code
Illustrative sample
Illustrative sample. Synthetic data, not from any employer or client.
Forty simulated paired readings. On average the sensor reads 0.34 m/s higher than the reference, and 95% of differences fall between -0.75 and 1.42 m/s. The gap widens at higher readings (slope 0.29), so one fixed correction would not be enough.
The method is the Bland-Altman plot [5]: plot each pair’s difference against its mean, rather than a correlation, which can look strong even when two devices disagree.
# Python. Full script on the Code and Notebooks page.
import numpy as np
rng = np.random.default_rng(2026) # fixed seed, reproducible
reference = rng.normal(7.0, 1.2, 40) # reference PWV, m/s
sensor = (reference + 0.25 + 0.08 * (reference - 7.0)
+ rng.normal(0, 0.45, 40)) # small bias and slope
mean_pair = (sensor + reference) / 2
diff = sensor - reference
bias = diff.mean()
sd = diff.std(ddof=1)
limits = (bias - 1.96 * sd, bias + 1.96 * sd)
slope, _ = np.polyfit(mean_pair, diff, 1) # does the gap grow?
print(f"Bias {bias:.2f}; limits {limits[0]:.2f} to {limits[1]:.2f}")
# Bias 0.34; limits -0.75 to 1.42
The finger result changed what we claimed. The paper concludes that PPG can capture acute change in the lower limb, and that the upper-limb method needs more work [1]. Stating that limit is what makes the rest of the claim usable.
References
Zieff G, Stone K, Paterson C, Fryer S, Diana J, Blackwell J, Meyer ML, Stoner L. Pulse-wave velocity assessments derived from a simple photoplethysmography device: agreement with a referent device. Front Cardiovasc Med. 2023;10:1108219. doi:10.3389/fcvm.2023.1108219
Luco AA, Gram M, Ramakers I, Ryan T, Enriquez O, Faulkner J, Stoner L, Rowlands DS. Skeletal muscle microvascular and respiratory responses to exercise in healthy and T2D adults: agreement and reproducibility study of NIRS technologies. Eur J Appl Physiol. 2026;126(9):4817-4831. doi:10.1007/s00421-026-06300-y
Stone K, Fryer S, Meyer ML, et al., Stoner L. The aortic-femoral arterial stiffness gradient: an atherosclerosis risk in communities (ARIC) study. J Hypertens. 2021;39(7):1370-1377. doi:10.1097/HJH.0000000000002808
Stone K, Fryer S, Faulkner J, et al., Stoner L. The aortic-femoral arterial stiffness gradient is blood pressure independent in older adults: the ARIC study. J Hypertens. 2021;39(12):2361-2369. doi:10.1097/HJH.0000000000002937
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307-310. PubMed 2868172